Date | Description | Person(s) | Hardware Group | Action performed | Attachment | Edit | Site | Task(?) |
2024-04-18 |
After coming home for dinner, no one on watch, so took it. Monitor was not running, so started it .... after ~10min things started, but monitor stopped itself, tried to restart, but it stopped again. The 'client_abort_acq' did not work, so stopped or_scheduler. Then tried to close the hatch and dome ... the hatch closed, but when the upper hatch closed I could see the lower hatch was not closed properly, despite Astelos reporting it closed ... so opened the uper/lower hatces agin... then the telescope slewed to a new objece (despite the monitor was not running!): ... then I closed the lower and uppper hatch, parked the telescope ... during which the 'both-limits-active' ocurred.. and then the telescope powered off.... Not sure what all the reasons for this are, but guessing perhaps the new(?) monitor. |
Frank Grundahl |
All |
see description above. |
None |
Edit |
Tenerife |
210 |
2024-04-05 |
Expected sudden rain at OT tonight and tomorrrow (only) |
Pere |
All |
Watchman will look closely |
None |
Edit |
Tenerife |
|
2024-04-02 |
A number of short power failures |
Mads |
All |
Cleared the error in Astelos (telescope control software) when errors occurred. Power cuts from 1 to around 10 seconds. |
Show |
Edit |
Tenerife |
|
2024-04-02 |
Power failures continued - last part of night lost |
Mads |
Telescope |
Short power failures continued during night and prevented reopening after weather improved. |
None |
Edit |
Tenerife |
|
2024-04-02 |
The dome did not close correctly in the morning |
Mads |
Dome |
Telescope and dome control system reported an error which after acknowledged the dome closed correctly. Next time Antonio on duty he could check the dome slit gear. Maybe it needs to be replaced? |
None |
Edit |
Tenerife |
209 |
2024-03-26 |
Half yearly regular RCD electrical testing |
Mads |
Electrical |
Stopped software and powered off tel-coms (t1, t2, fut), motors, detectors and main server. |
None |
Edit |
MtKent |
|
2024-03-20 |
washed M1 miorror with distilled water and dried with microfiber cloth |
Antonio, Maintanense grup |
Telescope |
washed M1 miorror |
None |
Edit |
Tenerife |
|
2024-03-12 |
T1 guider with new USB extender reports error (ASI_ERROR_TIMEOUT) |
Mads |
T1 guider |
The camera is detected on the server but will not acquire images. I suggest a hard reset (power cycle). Is the USB extender power connected to the PDU outlet number 6? |
None |
Edit |
MtKent |
208 |
2024-03-04 |
Filterwheel was not placed correctly in front of slit guider - perhaps causing flux loss |
Mads |
Spectrograph |
Homed the filterwheel which fixed the issue. |
None |
Edit |
Tenerife |
|
2024-02-09 |
No connection to anything in FUT Dome. Perhaps no power? Seem to recall something about a breaker switch separate for the FUT Dome in previous similar situation. I send this on behalf of Mads, who is going on vacation. |
Karsten |
All |
|
None |
Edit |
MtKent |
207 |
2024-02-07 |
Thorium Argon lamp change |
Antonio P. |
Spectrograph |
Thorium Argon lamp change |
Show |
Edit |
Tenerife |
|
2024-02-06 |
Most of the night lost because of hanging QHY600 CMOS detector |
Mads |
Detector |
Restared the daemon of the detector as a one time fix. Long term solution is being investigated. |
None |
Edit |
Tenerife |
206 |
2024-02-05 |
Guide camera on telescope 1 is not responding correctly. |
Mads |
Guiders |
Tried to power cycle outlet 6 on the T1 PDU in the dome. This should be the USB booster for the ZWO camera. It did not help. Perhaps a physical unplug - wait 30 seconds - and replug will work. |
None |
Edit |
MtKent |
205 |
2024-02-03 |
Blowing M1/M3 mirrors |
Antonio P. |
Telescope |
spots not removed. I'll contac Maintenace OT for aproper washing (TBC) |
None |
Edit |
Tenerife |
|
2024-02-03 |
CHECKED IN SITU THE THRAR CONTROLLER UNIT |
Antonio P. and Pere (remotely) |
Spectrograph |
Confirm that unit itself OK but always red led ON, "Faulty". Power cycle and other actions but no changes at all |
Show |
Edit |
Tenerife |
|
2024-02-01 |
No connection to FUT CMOS detector |
Mads |
Detector |
The USB connection does not show up on the machine on which the software is running. Maybe a loose USB connector? |
None |
Edit |
MtKent |
204 |
2024-02-01 |
Thorium Argon lamp controller reporting faults |
Mads |
Lamp controller |
Tried to power cycle the controller but the issue is still persistant. Local investigations of the unit is needed. Bottom unit on the shelf at the back of the container. |
None |
Edit |
Tenerife |
203 |
2024-01-29 |
Header entries for slit 6 calibration files was set to 0 ... so pipeline fails. |
Frank |
Spectrograph Detector |
None, just something to keep an eye on. The slit id was correct for the rest of the night. |
None |
Edit |
Tenerife |
|
2024-01-29 |
There seem to be a problem with the slit-guider. |
Frank Grundahl |
Slit-guider |
Tried to stop/start the slit guider on 192.168.67.1 ... looking in the log resulted in error message - see attachment. Activate full shutdown as we cannot observe more tonight. |
Show |
Edit |
Tenerife |
202 |
2024-01-12 |
Moved the control of the slit guider to the new detector PC in stead of the old kvmtenerife |
Mads |
Guider |
Installed the guide software on the new PC and moved the USB cable to connect to this machine. |
None |
Edit |
Tenerife |
|
2024-01-11 |
The spectrograph detector was changed from Andor Ikon-L CCD to QHY600 Pro CMOS |
Mads |
Spectrograph |
Change them |
None |
Edit |
Tenerife |
|
2024-01-10 |
Greased PST motor spindles |
Mads |
Motors |
Used grease from the Pyramid and added this gently to the spindle rods and afterwards moved the moving part back and forth a number of times. |
None |
Edit |
Tenerife |
|
2023-12-12 |
Observations did not start after sunset, and the weather is fine. |
Fran |
Servers ? |
PROBLEM is gone/solved -- PERHAPS there were no objects available for the first ~30min ... seem ok now. Tried to abort or_req 93185 ... as that was in the queue, but only for execution after midnight. Other time critical ORs are inserted for after 22:00 UT, perhaps that is affecting something ? Comment fom Mads: The conductor had crashed eariler that day. |
None |
Edit |
Tenerife |
201 |
2023-12-12 |
Temperature in spec. room too high |
Frank Grundahl |
Spectrograph room |
Emailed Jack who will contact building services. |
None |
Edit |
MtKent |
200 |
2023-12-11 |
Powered off tel-coms, detector, motors and server |
Mads |
All |
Preventive due to power cut tomorrow for UPS work |
None |
Edit |
MtKent |
|
2023-12-09 |
No weather station data since 07:51 UTC |
Mads |
Weather station |
Both data streams from the Boltwood and the Davis sensor stopped at 07:51 UTC. Guess is that the MINERVA/USQ script fails? |
None |
Edit |
MtKent |
199 |
2023-12-05 |
Boltwood sensor not reporting correct sky temperatures (cloud condition) |
Mads |
Weather station |
Changed limits in the software to accomodate for the bad readings - readings seemed to be offset. Maybe dirty sensor head? |
None |
Edit |
MtKent |
|
2023-12-04 |
No connection to the FUT dome and everything inside |
Mads |
Electrical / power |
Checked fibre connection but looks very similar to the issue last week with a flipped circuit breaker. Is something getting wet inside the FUT dome? |
None |
Edit |
MtKent |
198 |
2023-12-02 |
Observations does not seem to be carried out - telescope can point, but does not seem to start observations |
Frank Grundahl |
Telescope |
Tried stop/start monitor and tcs_comm demon - nothing seem to change. Cancelled a lot of ORs in 'wait' state. No effect. Stopped the monitor, closed dome, parked telescope and povered off. Closed the side-ports. |
None |
Edit |
Tenerife |
197 |
2023-11-28 |
Replacement of the ThAr lamp |
Frank Grundahl |
Spectrograph |
Duncan and Jack exchanged the SONG ThAr lamp which was throughly 'worn' ... see attached image. Flux checked and is fine after replacement. A lamp from MINERVA was used, as the spare SONG lamp does not seem to fit the mount. |
Show |
Edit |
MtKent |
|
2023-11-23 |
No connection to the FUT dome, telescope, computers etc |
Mads |
All |
Checked if the fiber connection needed to be reactivated but that did not help. Looks like a power loss or some other kind of bad network connection. It is only affecting the FUT. |
None |
Edit |
MtKent |
196 |
2023-11-18 |
The pointing model was not so great and targets were some times not acquired. |
Mads |
Telescope |
Made a new pointing model for coude: RMS = 1.03" with 76 stars |
None |
Edit |
Tenerife |
|
2023-11-13 |
Changed batteries on main SONG UPS |
Antonio, Pere, Mads |
Servers |
While servers were powered off due to planned power cut the battery for the UPS was changed and tested (12 minutes). |
None |
Edit |
Tenerife |
|
2023-11-09 |
No connection to dome 2 and the FUT dome |
Mads |
Network |
Since we have connection to SONG dome 1 this is again most likely an issue with the fibre connectors for Dome 2 and the FUT dome. Unplugging and replugging could be the fix. |
None |
Edit |
MtKent |
195 |
2023-10-26 |
Observations stopped due to 'some' critical condition. It seems that the status from PWI for alt and az does not match the OBS pages, not the web cameras. |
Frank Grundahl |
Telescope |
Tried to stop/start the monitor this morning (could not locate problems in the logs). It is cloudy, so perhaps related to guiders not being able to find the guide stars. I have put this under 'important', but it can wait til tomorrow. |
None |
Edit |
MtKent |
194 |
2023-10-25 |
The aircondition in the spectrograph room seem to go 'bananas' resulting in a dramatic temperature increase in the room |
Frank Grundahl |
Spectrograph |
None - created this post -- is this perhaps the annual adjustment of the system when switching to 'summer-mode' ? (done by USQ building services..?). |
Show |
Edit |
MtKent |
193 |
2023-10-18 |
Observations stopped due to 'some' critical condition |
Frank Grundahl |
? |
Total shutdown, no reason to stay open - I cannot find the cause for this in the logs - guessing it might be the guide camera for T1 .... |
None |
Edit |
MtKent |
192 |
2023-10-06 |
Pupil guider seems broken |
Mads |
Pupil guider |
Tried to restart and power cycle it but the software reports an error on the readout from the camera. It has been disabled in the observing scripts. It was not actively used for anything else than checking. |
None |
Edit |
Tenerife |
|
2023-10-06 |
Network issue after electrical tests |
Mads, Jack |
Many |
Rebooted several things, replugged many things. Nothing worked or works! |
None |
Edit |
MtKent |
191 |
2023-10-05 |
Reoccurring electrical instalations check - scheduled power failures |
Mads |
Many different |
Powered off computers and other vulnerable electical hardware - will restart all tomorrow (UTC) / later (local time). Jack checks fibre connectors and sets shutters to remote in domes. |
None |
Edit |
MtKent |
190 |
2023-10-04 |
Power failure from 16:50 UTC and the whole night |
Mads, Pere, Antonio |
All |
The power cable from Cielo Nocturno was bad. Antonio switched back to old power cable. |
None |
Edit |
Tenerife |
|
2023-10-04 |
Power failure in FUT dome - now back online |
Mads |
FUT electronics |
The components had power cycled in the FUT dome. So a temporary power failure apparently happened. All seems to be working again. |
None |
Edit |
MtKent |
|
2023-10-02 |
There seem to be no connection to the site - sstenerife and the 'obs' pages cannot be contacted. When trying to get the 'Observer' page up an error messageConnection timed out Is the server running on host "dbhub.sstenerife.prv" occurs which says: |
Frank Grundahl |
Servers -- sonnection |
None - I do not know how to fix the problem. |
None |
Edit |
Tenerife |
189 |
2023-10-01 |
No connection to SONG and FUT domes (tel-coms, dome, telescope, webcams etc) |
Mads |
Dome connection |
Could not connect to anything in the domes. Indicates issue with network connection - the fibres. Suggested on site action: Unplug/replug fibre connectors at server side. If not working reset (power cycle) fibre unit in the domes. |
None |
Edit |
MtKent |
188 |
2023-09-14 |
Weather data not updated for about 2 hours now |
Mads |
Weather station |
It seems the weather data from MINERVA is not being updated |
None |
Edit |
MtKent |
187 |
2023-09-10 |
Slit guide camera not detectable (usb) |
Mads |
Slit guider |
Does not show up at all on server. Cables and plugs needs to be checked |
None |
Edit |
Tenerife |
186 |
2003-02-24 |
Checks ThrAr |
Antonio P., Pere |
Spectrograph |
Check the Controller Unit.. "Fault" red led always on.. power recycle, etc... No changes |
None |
Edit |
Tenerife |
|
1970-01-01 |
washed M1 miorror with distilled water and dried with microfiber cloth |
Antonio, Mantenamse grup |
Telescope |
washed M1 miorror |
Show |
Edit |
Tenerife |
|
1970-01-01 |
|
|
|
revision and cleaning of dust in container |
None |
Edit |
Tenerife |
|
1970-01-01 |
washed M1 miorror with distilled water and dried with microfiber cloth |
Antonio, Mantenamse grup |
Telescope |
washed M1 miorror |
Show |
Edit |
Tenerife |
|