SONG technical management tasks page

Update status of task:
Description of task:The calibration fibre for SONG telescope 1 is misaligned with the science fibre. See attachment

Updated by
(Re)assign to
New status
Comment
Expected handled on Expected to be handled on? [YYYY-MM-DD]


Old comments:

2025-06-23 10:49: [Mads]: Light from calibration unit was checked on the guide unit in the nasmtyh of T1. Afterwards light was send back through the science fibre to determine the location on the guider. These two do not align and the calibration fibre should be adjusted on site to hit the science fibre better.