SONG technical management tasks page
Details of SONG task
| Insertion date | Task | Responsible | Level |
| 2025-07-29 |
No calibration light through fibres for quite some time |
Mads |
Important |
| Handled date | Comment | Handled by | Task ID |
| 2025-11-04 |
2025-11-04 10:49: [Mads]: Many things performed... in the end it was the position of the iodine cell that was not correct and obstructing the light path
2025-07-29 10:33: [Mads]: Checked if light from ThAr and Flat came to guide unit in nasmyth which it did not - most likely the shutters in the guide unit are not activated in remote mode. @Duncan: could you please check and activate them next time at MtKent? |
Mads |
266 |
|