SONG technical management tasks page

Details of SONG task
Insertion dateTaskResponsibleLevel
2023-12-03 Observations does not seem to be carried out - telescope can point, but does not seem to start observations Frank Grundahl Important

Handled dateCommentHandled byTask ID
2023-12-04 2023-12-04 07:51: [Mads]: Slit guide camera needed to have a USB reset.

2023-12-03 00:15: [Frank Grundahl]: Tried stop/start monitor and tcs_comm demon - nothing seem to change. Cancelled a lot of ORs in 'wait' state. No effect. Stopped the monitor, closed dome, parked telescope and povered off. Closed the side-ports.
Mads 197