SONG technical management tasks page
Details of SONG task
Insertion date | Task | Responsible | Level |
2022-08-16 |
The spectrograph only has calibration light for one fiber. Cause unknown, but first test is to check whether the calibration shutters are in 'remote' mode. Please check this for both domes, and make sure both are enabled, in case they are not. We still get starlight from both telescopes, so it's strictly a 'calibration issue'. |
Frank Grundahl |
Important |
Handled date | Comment | Handled by | Task ID |
2023-01-17 |
2023-01-17 06:00: [Jack]: The shutter controller was not in remote mode.
2022-08-18 03:28: [Jack]: The shutter controller on telescope 1 was off, we've turned it back on and set it to remote mode. Let me know if that fixes the issue.
2022-08-17 04:19: [Jack]: I'll reset the shutters in the dome to the remote setting tomorrow when Duncan and I are on site. |
Jack |
142 |
|