SONG technical management tasks page
Details of SONG task
Insertion date | Task | Responsible | Level |
2022-03-21 |
No connection from server to CMOS detector in the spectrograph. On-site check of USB connection needed. |
Mads |
Critical |
Handled date | Comment | Handled by | Task ID |
2022-03-24 |
2022-03-24 09:19: [Duncan]: The usb connection to the camera was pulled out when removing the thar power supply
2022-03-22 07:15: I'll speak with Duncan, and organise a time to investigate in case we need to check the connection to the camera inside the enclosure.
Power cycled the Kepler4040 in the spectrograph and the USB hub in which the detector -might- go through. Did not help. The detector does not show up using 'lsusb'. |
Duncan |
115 |
|